Reference
SCPI Command :
SENSe<CH>:[POWer]:SWEep:TIME:[SENSor]:PULSe:THReshold:POWer:REFerence
- class ReferenceCls[source]
Reference commands group definition. 1 total commands, 0 Subgroups, 1 group commands
- get(channel=Channel.Default) float[source]
# SCPI: SENSe<CH>:[POWer]:SWEep:TIME:[SENSor]:PULSe:THReshold:POWer:REFerence value: float = driver.sense.power.sweep.time.sensor.pulse.threshold.power.reference.get(channel = repcap.Channel.Default)
Sets the medial reference level in terms of percentage of the overall pulse level (power or voltage related) . This level is used to define pulse width and pulse period. Note: The command is only available in time measurement mode and with R&S NRPZ81 power sensors.
- param channel:
optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Sense’)
- return:
reference: float Range: 0.0 to 100.0
- set(reference: float, channel=Channel.Default) None[source]
# SCPI: SENSe<CH>:[POWer]:SWEep:TIME:[SENSor]:PULSe:THReshold:POWer:REFerence driver.sense.power.sweep.time.sensor.pulse.threshold.power.reference.set(reference = 1.0, channel = repcap.Channel.Default)
Sets the medial reference level in terms of percentage of the overall pulse level (power or voltage related) . This level is used to define pulse width and pulse period. Note: The command is only available in time measurement mode and with R&S NRPZ81 power sensors.
- param reference:
float Range: 0.0 to 100.0
- param channel:
optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Sense’)