Reference

SCPI Command :

SENSe<CH>:[POWer]:SWEep:TIME:[SENSor]:PULSe:THReshold:POWer:REFerence
class ReferenceCls[source]

Reference commands group definition. 1 total commands, 0 Subgroups, 1 group commands

get(channel=Channel.Default) float[source]
# SCPI: SENSe<CH>:[POWer]:SWEep:TIME:[SENSor]:PULSe:THReshold:POWer:REFerence
value: float = driver.sense.power.sweep.time.sensor.pulse.threshold.power.reference.get(channel = repcap.Channel.Default)

Sets the medial reference level in terms of percentage of the overall pulse level (power or voltage related) . This level is used to define pulse width and pulse period. Note: The command is only available in time measurement mode and with R&S NRPZ81 power sensors.

param channel:

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Sense’)

return:

reference: float Range: 0.0 to 100.0

set(reference: float, channel=Channel.Default) None[source]
# SCPI: SENSe<CH>:[POWer]:SWEep:TIME:[SENSor]:PULSe:THReshold:POWer:REFerence
driver.sense.power.sweep.time.sensor.pulse.threshold.power.reference.set(reference = 1.0, channel = repcap.Channel.Default)

Sets the medial reference level in terms of percentage of the overall pulse level (power or voltage related) . This level is used to define pulse width and pulse period. Note: The command is only available in time measurement mode and with R&S NRPZ81 power sensors.

param reference:

float Range: 0.0 to 100.0

param channel:

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Sense’)